Whether your testing is for R&D, clinical trials or ongoing stability, “TEMPO” stability chamber provide a stable temperature & humidity condition environment for worry-free operations with a control system that is easy to operate. “TEMPO” make stability chamber is designed as per ICH guidelines and are manufactured to comply with requirements in the industries. Stability testing is an integral part of formulation development. It helps to generate information, which permits well-considered proposals to be made for the shelf life of drug substances and products and recommended storage condition.

The purpose of stability testing is to provide evidence on how the quality of a drug substance or drug product varies with time under the influence of a variety of environmental factors such as temperature, humidity, and light, and to establish a re-test period for the drug substance or product under recommended storage conditions. Hence, pharmaceutical stability chamber serves as an important quality attribute for the product.

Temperature range – 20°C to 60°C
Temperature accuracy – ±0.2°C
Temperature uniformity – ±1°C
Humidity Range – 40% Rh to 90% Rh
Humidity Accuracy – ±2%Rh
Humidity Uniformity – ±3%Rh

Construction
Double walled construction with backside triple walled 3” thick PUF insulation ensures stable temperature with humidity & reduced energy consumption. Motorized Blower is fitted at the backside of the chamber.

The unit is provided with an adjustable stainless steel rod mesh tray. The equipment is mounted on caster wheels for easy movement.

Technical Specification
Controlling System
Two separate imported microprocessors based autotune PID controller with CE mark & dual display of set value & process value for temperature & humidity with direct %RH.

Air Circulation
The motorized blower at the backside of the chamber develops a unique air ow system which ensures max. uniform temperature & humidity condition inside the chamber.

Heating System
High-quality U shape S.S. tubular heater is used for better heating.

Humidity System
Steam injection system (Non-condensing type) & reservoir tank with water level arrangement with indicator.

Refrigeration System
CFC Free hermetically sealed Emerson Copeland make compressor for better cooling with R134 A (Eco Friendly) refrigerant with a time delay to safeguard compressor system.

Features
50 mm validation porthole with silicon rubber seal to insert sensors for validation. Full view observation Acrylic door with a gasket to observe the sample inside the chamber. Interior illumination for working area.
Specially designed stainless steel rod trays ensure uniform temperature distribution.

Safety Features
Additional safety thermostat to cut of dry heater supply in case temperature of overshoot. Float switch to cut off the wet heater supply in case of low water level. Audio & visual alarm. Compressor ON delay timer (2 min.) to safeguard the compressor. Inbuilt overload protector provided for a hermetically sealed compressor.